Screening for Chronic Obstructive Pulmonary Disease: Validity and Reliability of a Portable Device in Non-Specialized Healthcare Settings.
Identificadores
Identificadores
Visualización ou descarga de ficheiros
Visualización ou descarga de ficheiros
Autor corporativo
Study group of Validity of COPD-6 in nonspecialized healthcare settingsData de publicación
2016Título da revista
PloS one
Tipo de contido
Artigo
DeCS
enfermedad pulmonar obstructiva crónica | hábito de fumarMeSH
Smoking | Pulmonary Disease, Chronic ObstructiveResumo
The underdiagnosis of chronic obstructive pulmonary disease (COPD) could be improved through screening using portable devices simpler than conventional spirometers in specific healthcare settings to reach a higher percentage of the at-risk population. This study was designed to assess the validity and reliability of the COPD-6 portable device to screen for COPD in non-specialized healthcare settings.
Prospective cohort study to validate a diagnostic test. Three cohorts were recruited: primary care (PC), emergency services (ES) and community pharmacies (CPh).
individuals with risk factors for COPD (>40 years, smoking >10 pack-years, with respiratory symptoms). The values measured using the COPD-6 were FEV1, FEV6 and the FEV1/FEV6 ratio. Subsequently, participants underwent conventional spirometry at hospital, using a post-bronchodilator FEV1/FVC value <0.7 as the gold standard criterion for the COPD diagnosis.
437 participants were included, 362 were valid for the analysis. COPD was diagnosed in 114 patients (31.5%). The area under the ROC curve for the COPD-6 for COPD screening was 0.8.The best cut-off point for the FEV1/FEV6 ratio was 0.8 (sensitivity, 92.1%) using spirometry with the bronchodilator test as the gold standard. There were practically no differences in the COPD-6 performance in the different settings and also regarding age, gender and smoking status.
The COPD-6 device is a valid tool for COPD screening in non-specialized healthcare settings. In this context, the best cut-off point for the FEV1/FEV6 ratio is 0.8.
O ítem ten asociados os seguintes ficheiros de licenza: