Mostrar el registro sencillo del ítem

dc.contributor.authorDurán Bobín, Olga
dc.contributor.authorCrespo Mancebo, Francisco 
dc.contributor.authorElices Teja, Juliana 
dc.contributor.authorGonzález Juanatey, Carlos 
dc.date.accessioned2024-01-02T10:02:01Z
dc.date.available2024-01-02T10:02:01Z
dc.date.issued2021
dc.identifier.issn2514-2119
dc.identifier.otherhttps://www.ncbi.nlm.nih.gov/pubmed/33629026es
dc.identifier.urihttp://hdl.handle.net/20.500.11940/18423
dc.description.abstractBACKGROUND: Syncope in a patient with a pacemaker is a serious event requiring urgent action to ascertain its cause. Around 5% of cases are due to a pacemaker system malfunction. CASE SUMMARY: An 82-year-old man underwent dual-chamber permanent pacemaker implantation due to intermittent high-degree atrioventricular block (AVB) in sinus rhythm. Nine months later, the patient reported episodes of syncope. The chest X-ray showed both leads to be at their expected positions. The electrocardiography (ECG) showed common atrial flutter. Ventricular capture during pacing in atrial demand pacing (AAI) mode confirmed cross-stimulation due to the switching of the atrial and ventricular leads at the pacemaker header. DISCUSSION: Cross-stimulation is a rare possibility in a differential diagnosis of causes of syncope. The diagnosis is frequently made during the procedure or a few hours later. The lack of symptoms during 9 months in this case was likely due to the patient having normal sinus rhythm with preserved AV conduction most of the time, as well as ventricular capture from the atrial lead related to non-sensed P waves. When atrial arrhythmias occurred, the sensing of the F waves inhibited ventricular pacing. In order to avoid this complication, in patients with intermittent bradycardia, pacing at a slightly higher heart rate during implantation of the device should be recommended to see the chamber paced with the surface ECG connected to the device interrogator. The ECG and electrogram (EGM) should correlate during device interrogation in order to identify this complication.).
dc.language.isoen
dc.rightsAtribución-NoComercial 4.0 Internacional
dc.rights.urihttp://creativecommons.org/licenses/by-nc/4.0/
dc.titleSyncope triggered by atrial flutter in a patient with a pacemaker: cross-stimulation-a case report
dc.typeJournal Articlees
dc.authorsophosDurán-Bobin, O.;Crespo-Mancebo, F.;Elices-Teja, J.;González-Juanatey, C.
dc.identifier.doi10.1093/ehjcr/ytab041
dc.identifier.pmid33629026
dc.identifier.sophos49589
dc.issue.number2
dc.journal.titleEuropean Heart Journal Case Reports
dc.organizationServizo Galego de Saúde::Áreas Sanitarias (A.S.)::Área Sanitaria de Lugo, Cervo e Monforte de lemos - Complexo Hospitalario Universitario Lucus Augusti::Cardioloxía
dc.page.initial41
dc.rights.accessRightsopenAccess
dc.subject.keywordCHULAes
dc.typefidesArtículo de Opinión (Editorial, Carta, …)es