Improved image quality in pinhole SPECT by accurate modeling of the point spread function in low magnification systems

Pino, F.; Roé, N.; Aguiar Fernández, Pablo; Falcon, C.; Ros, D.; Pavía, J
Thumbnail
Estadísticas
Estadísticas
Ver Estadísticas de uso
Identificadores
Identificadores
URI: http://hdl.handle.net/20.500.11940/8326
PMID: 25652484
DOI: 10.1118/1.4905157
ISSN: 0094-2405
Registro completo
Servicios
Servicios
RISMendeleyLinksolver
Visualización o descarga de ficheros
Visualización o descarga de ficheros
Med Phys . 2015 Feb;42(2):703-14. doi: 10.1118/1.4905157. (6.145Mb)
Fecha de publicación
2015
Título de revista
MEDICAL PHYSICS
Tipo de contenido
Artigo
MeSH
Algorithms | Animals | Gamma Rays | Image Enhancement | Mice | Models, Theoretical | Phantoms, Imaging | Quality Control | Signal-To-Noise Ratio | Tomography, Emission-Computed, Single-Photon
Resumen
PURPOSE: Single photon emission computed tomography (SPECT) has become an important noninvasive imaging technique in small-animal research. Due to the high resolution required in small-animal SPECT systems, the spatially variant system response needs to be included in the reconstruction algorithm. Accurate modeling of the system response should result in a major improvement in the quality of reconstructed images. The aim of this study was to quantitatively assess the impact that an accurate modeling of spatially variant collimator/detector response has on image-quality parameters, using a low magnification SPECT system equipped with a pinhole collimator and a small gamma camera. METHODS: Three methods were used to model the point spread function (PSF). For the first, only the geometrical pinhole aperture was included in the PSF. For the second, the septal penetration through the pinhole collimator was added. In the third method, the measured intrinsic detector response was incorporated. Tomographic spatial resolution was evaluated and contrast, recovery coefficients, contrast-to-noise ratio, and noise were quantified using a custom-built NEMA NU 4-2008 image-quality phantom. RESULTS: A high correlation was found between the experimental data corresponding to intrinsic detector response and the fitted values obtained by means of an asymmetric Gaussian distribution. For all PSF models, resolution improved as the distance from the point source to the center of the field of view increased and when the acquisition radius diminished. An improvement of resolution was observed after a minimum of five iterations when the PSF modeling included more corrections. Contrast, recovery coefficients, and contrast-to-noise ratio were better for the same level of noise in the image when more accurate models were included. Ring-type artifacts were observed when the number of iterations exceeded 12. CONCLUSIONS: Accurate modeling of the PSF improves resolution, contrast, and recovery coefficients in the reconstructed images. To avoid the appearance of ring-type artifacts, the number of iterations should be limited. In low magnification systems, the intrinsic detector PSF plays a major role in improvement of the image-quality parameters.

Navega

Todo RUNAColeccionesCentrosAutoresTítulosDeCSMeSHCIETipos de contenidosEsta colecciónCentrosAutoresTítulosDeCSMeSHCIETipos de contenidos

Estadísticas

Ver Estadísticas de uso

DE INTERÉS

Sobre Acceso AbiertoDerechos de autor
TwitterRSS
Xunta de Galicia
© Xunta de Galicia. Información mantida e publicada na internet pola Consellería de Sanidade o Servizo Galego de Saúde
Aviso legal | RSS
Galicia