Geant4-GATE Simulation of a Large Plastic Scintillator for Muon Radiography
Identificadores
Identificadores
Visualización ou descarga de ficheiros
Visualización ou descarga de ficheiros
Data de publicación
2015Título da revista
IEEE TRANSACTIONS ON NUCLEAR SCIENCE
Tipo de contido
Artigo
MeSH
Cosmic rays | GATE | Geant4Resumo
Envisaging the possibility of using large-area plastic scintillator slabs as robust detectors for high spatial resolution muon radiography, and prior to prototype development, we study expected basic performance by Monte Carlo simulation. We present preliminary results for a scalable square footprint detector unit of similar to 1 m(2), defining a representative simulation model volume of 50 cm x 50 cm, with reflective surfaces and a light readout by direct coupling of 4 small PMTs (in a square arrangement) at one face of the scintillator slab. Light detection efficiency is calculated for several light collection configurations, considering different values of surface roughness, reflectivity, optical coupling index and scintillator thickness. Values maximizing photon detection have been identified. The light response function of 2.5-3.5 cm diameter PMTs for the proposed configuration has been determined. A detector intrinsic spatial resolution of the order of 1 cm is estimated for muon interactions at the center region of the detector module, using a simple centroid positioning algorithm (Anger logic).